Measurements and Results
R&S
®
FSW
292
User Manual 1173.9411.02 ─ 43
7.7.6
Reference: ASCII Export File Format (Spurious)
The file has a header containing important parameters for scaling, several data sec-
tions containing the sweep settings per range, and a data section containing the peak
list.
The header data is made up of three columns, separated by ';', with the syntax:
Parameter name; numeric value; basic unit
File contents
Explanation
File header
Type;FSW-26;
Model
Version;1.00;
Firmware version
Date;31.Mar 11;
Storage date of data set
Mode;ANALYZER; SPURIOUS;
Operating mode and measurement function
Center Freq;13250000000.000000;Hz
X-axis settings
Freq Offset;0.000000;Hz
Span;26499982000.000000;Hz
x-Axis;LIN;
Start;9000.000000;Hz
Stop;8000000000.000000;Hz
Level Offset;0.000000;dB
Y-axis settings
Ref Position;100.000000;%
y-Axis;LOG;
Level Range;100.000000;dB
Trace settings
Trace Mode;CLR/WRITE;
Sweep Count;1;
TRACE 1:
Trace Mode;CLR/WRITE;
x-Unit;Hz;
y-Unit;dBm;
List evaluation settings
Margin;6.000000;s
Peak List margin
PeaksPerRange;25;
Max. number of peaks per range to be detected
Values;3;
Number of detected peaks
Spurious Emissions Measurement