Remote Commands
R&S
®
FSW
1074
User Manual 1173.9411.02 ─ 43
Suffix:
<ip>
.
1 | 2
For R&S
FSW85 models with two RF input connectors:
1: Input 1 (1
mm [RF Input] connector)
2: Input 2 (1.85
mm [RF2 Input] connector)
For all other models:
irrelevant
Parameters:
<State>
ON | 1
User port is switched to INPut
OFF | 0
User port is switched to OUTPut
*RST:
1
INPut<ip>:UPORt[:VALue]
This command queries the control lines of the user ports.
For details see
Suffix:
<ip>
.
1 | 2
For R&S
FSW85 models with two RF input connectors:
1: Input 1 (1
mm [RF Input] connector)
2: Input 2 (1.85
mm [RF2 Input] connector)
For all other models:
irrelevant
Return values:
<Level>
bit values in hexadecimal format
TTL type voltage levels (max. 5V)
Range:
#B00000000 to #B00111111
Example:
INP:UPOR?
//Result: #B00100100
Pins 5 and 7 are active.
14.7.6.2
Using External Mixers
The commands required to work with external mixers in a remote environment are
described here. Note that these commands require the R&S
FSW-B21 option to be
installed and an external mixer to be connected to the front panel of the R&S
FSW.
In MSRA/MSRT mode, external mixers are not supported.
For details on working with external mixers see
Chapter 8.2.5.1, "Basics on External
Setting Basic Measurement Parameters