Remote Commands
R&S
®
FSW
1054
User Manual 1173.9411.02 ─ 43
EXTernal
Trigger signal from the "Trigger Input" connector.
If the optional 2
GHz bandwidth extension (B2000/B5000) is
installed and active, this parameter activates the "Ch3" input
connector on the oscilloscope. Then the R&S
FSW triggers
when the signal fed into the "Ch3" input connector on the oscillo-
scope meets or exceeds the specified trigger level.
Note:
In previous firmware versions, the external trigger was
connected to the "Ch2" input on the oscilloscope. As of firmware
version R&S
FSW 2.30, the
"Ch3"
input on the oscilloscope
must be used!
EXT2
Trigger signal from the "Trigger Input/Output" connector.
For R&S
FSW85 models, Trigger 2 is not available due to the
second RF input connector on the front panel. The trigger signal
is taken from the "Trigger Input/Output" connector on the rear
panel.
Note: Connector must be configured for "Input".
EXT3
Trigger signal from the "TRIGGER 3 INPUT/ OUTPUT" connec-
tor.
Note: Connector must be configured for "Input".
RFPower
First intermediate frequency
(Frequency and time domain measurements only.)
IFPower
Second intermediate frequency
TIME
Time interval
VIDeo
Video mode is available in the time domain and only in the
Spectrum application.
PSEN
External power sensor
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Setting Basic Measurement Parameters