Remote Commands
R&S
®
FSW
1205
User Manual 1173.9411.02 ─ 43
on page 1203 has been performed and thus does not begin by presetting the
instrument.
//------------- Configuring the phase noise marker -------------------
DET SAMP
//Switches to Sample detector
CALC:MARK1 ON
//Activates marker1 and sets it to the maximum power level
CALC:DELT:FUNC:PNO ON
//Activates phase noise marker function
CALC:DELT1 ON
CALC:DELT1:X 100kHz
CALC:DELT2 ON
CALC:DELT2:X 500kHz
CALC:DELT3 ON
CALC:DELT3:X 1MHz
CALC:DELT4 ON
CALC:DELT4:X 1.5MHz
//Activates the phase noise measurement function for offsets 100kHz/500kHz/1MHz/1.5MHz.
BAND:VID?
//Queries the used VBW (= 0.1*RBW)
//-------- Querying the phase noise results ------------------------------
CALC:DELT1:FUNC:PNO:RES?
CALC:DELT2:FUNC:PNO:RES?
CALC:DELT3:FUNC:PNO:RES?
CALC:DELT4:FUNC:PNO:RES?
//Queries the difference in level between the peak and the noise power density
//measured at the deltamarkers, referred to the carrier power level (dBc)
Example: Measuring the Power in a Channel Using Band Power Markers
This example demonstrates how to measure the power in a specific channel or band
using markers in a basic spectrum measurement in a remote environment. It assumes
that the basic frequency sweep described in
"Basic Frequency Sweep Measurement
on page 1203 has been performed and thus does not
begin by presetting the instrument.
//------------- Configuring the band power marker -------------------
CALC:MARK1 ON
//Activates marker1 and sets it to the maximum power level
Analyzing Measurements (Basics)