Remote Commands
R&S
®
FSW
882
User Manual 1173.9411.02 ─ 43
[SENSe:]POWer:ACHannel:BWIDth:GAP<gap>:MANual:UPPer
<SBGaps>,
<Bandwidth>
[SENSe:]POWer:ACHannel:BWIDth:GAP<gap>:MANual:UPPer?
<SBGaps>
[SENSe:]POWer:ACHannel:BANDwidth:GAP<gap>:MANual:UPPer
<SBGaps>,
<Bandwidth>
[SENSe:]POWer:ACHannel:BANDwidth:GAP<gap>:MANual:UPPer?
<SBGaps>
Defines the bandwidth of the upper gap channel in the specified gap.
Suffix:
<gap>
.
1 | 2
Gap channel number
Parameters:
<Bandwidth>
*RST:
3.84 MHz
Default unit: HZ
Parameters for setting and query:
<SBGaps>
AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
POW:ACH:BAND:GAP:MAN:UPP BC,5MHz
Example:
See
"Example: Configuring and Performing an Asymmetrical
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer[:STATe]:GAP<gap>:MANual:LOWer
<SBGaps>,
<State>
[SENSe:]POWer:ACHannel:FILTer[:STATe]:GAP<gap>:MANual:LOWer?
<SBGaps>
This command turns the weighting filter for the specified lower gap channel on and off.
Suffix:
<gap>
.
1 | 2
Gap channel number
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Parameters for setting and query:
<SBGaps>
AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Configuring and Performing Measurements