Remote Commands
R&S
®
FSW
1332
User Manual 1173.9411.02 ─ 43
SYSTem:IFGain:MODE
<Mode>
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz and is
only available if a HP language is selected using
Parameters:
<Mode>
NORMal | PULSe
NORMal
Optimized for high dynamic range, overload limit is close to ref-
erence level.
PULSe
Optimized for pulsed signals, overload limit up to 10 dB above
reference level.
*RST:
NORM
Example:
SYST:IFG:MODE PULS
Manual operation:
See
SYSTem:HPADditional
<State>
Allows the use of HP commands
in addition to
SCPI commands for R&S
Parameters:
<State>
ON | OFF | 1 | 0
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Manual operation:
See
See
SYSTem:LANGuage
<Language>
This command selects the system language.
For details see
Chapter 13.2, "GPIB Languages"
Note
:This command is also used to emulate previous R&S signal and spectrum ana-
lyzers, making the
SYST:COMP
command obsolete.
Note:
For PSA89600 emulation, the option is indicated as "B7J" for the
*OPT?
query
("B7J, 140" if
is activated).
Note:
For R&S FSP/FSQ/FSU emulation, HP commands are not automatically also
allowed. In this case, set
ON
.
Emulating Other Instruments' Commands