Data Management
R&S
®
FSW
633
User Manual 1173.9411.02 ─ 43
Selecting Storage Location - Drive/ Path/ Files
...........................................................633
Startup Recall
Activates or deactivates the startup recall function. If activated, the settings stored in
the selected file are loaded each time the instrument is started or preset. If deactivated,
the default settings are loaded.
Note that only
instrument
settings files can be selected for the startup recall function,
not "Channel" files.
Remote command:
Selecting Storage Location - Drive/ Path/ Files
Select the storage location of the file on the instrument or an external drive.
The default storage location for the SEM settings files is:
C:\Program Files (x86)\Rohde-Schwarz\FSW\<version>\sem_std
.
Note:
Saving instrument settings in secure user mode.
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
Storing and Recalling Instrument Settings and Measurement Data