Common Analysis and Display Functions
R&S
®
FSW
573
User Manual 1173.9411.02 ─ 43
How to move the limit line vertically or horizontally
A configured limit line can easily be moved vertically or horizontally. Thus, a new limit
line can be easily generated based upon an existing limit line which has been shifted
horizontally.
1. In the "Line Config" dialog box, select the limit line.
2. To shift the complete limit line parallel in the horizontal direction, select the "X-Off-
set" button and enter an offset value.
To shift the complete limit line parallel in the vertical direction, select the "Y-Offset"
button and enter an offset value.
3. To shift the individual data points of a limit line by a fixed value (all at once):
a) Select the "Edit" button.
b) In the "Edit Limit Line" dialog box, select the "Shift x" or "Shift y" button and
define the shift width.
c) Save the shifted data points by selecting the "Save" button.
If activated, the limit line is shifted in the diagram.
How to export a limit line
Limit line configurations can be stored to an ASCII file for evaluation in other programs
or to be imported later for other measurements.
1. In the "Line Config" dialog box, select the limit line.
2. Select the "New" or "Edit" button.
3. Define the limit line as described in
"How to configure a new limit line"
4. Select "Export" to save the configuration to a file.
You are asked whether you would like to save the configuration internally on the
R&S
FSW first.
5. Select a file name and location for the limit line.
6. Select the decimal separator to be used in the file.
7. Select "Save".
The limit line is stored to a file with the specified name and the extension
.CSV
.
For details on the file format see
Chapter 9.4.2.4, "Reference: Limit Line File For-
How to import a limit line
Limit line configurations that are stored in an ASCII file and contain a minimum of
required data can be imported to the R&S
FSW.
For details on the required file format see
Chapter 9.4.2.4, "Reference: Limit Line File
1. In the "Line Config" dialog box, select the limit line.
Display and Limit Lines