Measurements and Results
R&S
®
FSW
345
User Manual 1173.9411.02 ─ 43
Remote command:
Final Test Detector
Defines the detector to be used for the final EMI test at the marker frequency.
This setting is only available if the EMI (R&S
FSW-K54) measurement option is instal-
led.
For details, see
Chapter 7.13.3.2, "Detectors and Dwell Time"
Note:
The trace detector configured in the trace settings is used for the initial peak
search only, see
Chapter 9.5.1.2, "Trace Settings"
"Off"
No final test is performed.
"PositivePeak"
Determines the maximum signal level that was detected during the
specified dwell time.
"Average"
Determines the average signal level of the samples that were collec-
ted during the specified dwell time.
"Quasi-Peak"
Determines the maximum signal level weighted to CISPR
16-1-1 that
was detected during the dwell time.
The "Quasi-Peak" detector is only available for the CISPR filter, and
not for an RBW of 1
MHz.
"CISPR Aver-
age"
Determines a weighted average signal level according to CISPR
16-1-1.
The average value according to CISPR 16-1-1 is the maximum value
detected while calculating the linear average value during the speci-
fied dwell time.
The "CISPR Average" detector is only available for the CISPR filter.
"RMS Aver-
age"
A combination of the RMS detector (for pulse repetition frequencies
above a corner frequency) and the Average detector (for pulse repeti-
tion frequencies below the corner frequency).
Lowpass filters of the second order determine the average value
(simulation of a mechanical pointer instrument).
The "RMS Average" detector is only available for the CISPR filter.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DETector
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DETector
Select Marker
The "Select Marker" function opens a dialog box to select and activate or deactivate
one or more markers quickly.
Electromagnetic Interference (EMI) Measurement