ABL90 FLEX reference manual
6. User-defined corrections
6-3
It should be noted that in order to define corrections:
Measurements should be taken on the ABL90 FLEX analyzer without user-
defined corrections, and on a single reference analyzer
A series of measurements that cover the entire measuring range should be
performed
The measurements should be made simultaneously on the ABL90 FLEX and
the reference analyzer, and samples must be handled correctly.
The slope and the offset must be calculated. The user may, for example,
make a linear correlation between the values measured on the ABL90 FLEX
analyzer and the reference analyzer, using the ABL90 FLEX analyzer as an
independent variable.
The user must verify the corrections that are entered.
Details of these procedures may be found in the section Test conditions in
chapter 7.
The slope/offset for each parameter are configured in General Setup >
Parameters and Input > Parameters. User-corrected values are marked with
a "*" after the result.
NOTICE: The user-defined corrections will be applied to Quality Control
measurements unless the "Apply parameter corrections to QC" option was
deactivated in Miscellaneous setup as described in the section Miscellaneous
setup in chapter 1.
For detailed instructions on how to enter user-defined corrections, refer to the
section Parameters and input setup in chapter 1.
Entering user-
defined
corrections
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...