10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-10
No.
Message
Interpretation
Operator action
508
Liquid transport
error during rinse
Liquid transport of Rinse
failed
- Check solution pack or
sensor cassette status
and replace, if
necessary.
Removal condition:
- Successful Rinse.
512
Temperature
error
The temperature was outside
the required range during
measurement or calibration.
All results are marked with
"?".
- Ensure that the
ambient temperature is
between 15 and 32 °C.
- If the analyzer has
recently performed a
cold start, wait for the
temperature error to
disappear.
- If the solution pack or
sensor cassette has
recently been replaced,
wait for the temperature
error to disappear.
- Shield analyzer from
direct sunlight or heat
sources.
- Contact Radiometer
service representative.
521
Inhomogeneous
sample
Air bubbles were detected in
the sample. Results may
have "?".
- Repeat the
measurement.
522
Calibration error One or more calibration
values are erroneous.
- Check for and remedy
any system messages.
- Repeat the calibration.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
Removal condition:
- Successful calibration
523
Calibration drift
out of range
Calibration drift exceeds
defined limits.
- Check for and remedy
any System Messages.
- Perform any pending
replacements including
electrodes.
- Check that electrodes
are properly installed.
- Verify that proper
solutions and gases are
used.
- Perform the Electrode
Troubleshooting
procedure.
Removal condition:
- Calibration drift within
defined limits.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...