ABL90 FLEX reference manual
Appendix – Quality control
I-13
This rule is a warning rule.
Measurement value is
outside the mean
2 SD.
Corrective action:
Perform another measurement on a new ampoule
of the same level. If the second result falls within
the control range, then the first result can be
attributed to normal statistical variation.
If the second result is outside the established
mean
2 SD, see section Quality control
evaluation
further in this appendix.
This rule is applicable to both manual and built-in QC.
This rule is a rejection rule.
Measurement value is
outside the mean
3 SD.
Corrective action:
Perform another measurement on a new ampoule
of the same level. If the second result falls within
the control range, then the first result can be
attributed to normal statistical variation.
If the second result is outside the established
mean
2 SD, see section Quality control
evaluation
further in this appendix.
This rule is applicable to both manual and built-in QC.
This rule is a rejection rule.
Two consecutive
measurements are
outside the mean
2 SD
on the same side of the
mean.
Corrective action:
Perform steps described in section Quality control
evaluation
further in this appendix.
This rule is applicable to both manual and built-in QC.
Rule 1
2s
Rule 1
3s
Rule 2
2s
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...