5. Sensors and optical system
ABL90 FLEX reference manual
5-48
The following parameters will not be calculated:
Parameter
Is not calculated if…
sO
2
,
FCOHb, FMetHb,
FHHb, FO
2
Hb
ctHb < 1 mmol/L
sO
2
ceHb = cHHb + cO
2
Hb < 0.75 mmol/L
ctBil
ctHb > 14.27 mmol/L
The following conditions are required to perform HbF suppression:
Parameter or Feature
Requirement
ctHb
> 5 mmol/L
FCOHb
< 20 %
FMetHb
< 10 %
HbF correction –
Enabled for levels
>20%
cO
2
HbF/
ctHb should be more than 0.2.
HbF correction –
Enabled for all levels
No lower limit value for
cO
2
HbF is required, i.e.
even adult blood samples will be corrected for HbF.
It may be of value when analyzing blood samples
from newborns who have received adult blood
transfusion. In these cases
FHbF can be lower than
20 % and significant deviations of oximetry
parameters and bilirubin can occur.
HbF correction –
Disabled
No HbF corrections made.
HbF suppression has
been activated
The
FHbF value is normally, but not always,
displayed by the analyzer.
Message "Oxi compensated for HbF" is displayed.
sO
2
< 50 %
Message "
FHbF measurement is not possible" is
displayed by the analyzer, if a HbF suppression has
been activated, and the
FHbF estimation from
cO
2
HbF is too uncertain.
Restrictions
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...