ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-31
No.
Message
Interpretation
Operator action
1125 An unregistered
and used sensor
was installed
during
replacement
This message is shown in the
Activity Log after a sensor
cassette replacement. It
informs that the sensor
cassette installed is already
used and no information
exists about the conditioning
hereof.
- No action required. For
information only.
1126 A registered
sensor had been
used before
installation
This message is shown in the
Activity Log after a sensor
cassette replacement. It
informs that the sensor
cassette installed has been
used before.
- No action required. For
information only.
1134 The chip
information for
the solution pack
cannot be read
or written
This message is shown in the
Activity Log after a failed
solution pack replacement.
- Reinstall the solution
pack.
- Restart the analyzer.
- Contact Radiometer
service representative.
1135 The solution pack
has been used
before
This message is shown in the
Activity Log after a failed
solution pack replacement.
- Reinstall the solution
pack.
1140 The solution pack
has used the
maximum
number of
measurements at
installation
This message is shown in the
Activity Log after a failed
solution pack replacement.
- Reinstall the solution
pack.
1142 The printer door
is open. Printing
not possible
Printer door open.
- Ensure that the printer
paper is properly
installed.
- Close the printer door.
1143 Internal printer is
offline. Printing
not possible
Printer hardware error.
- Ensure that the printer
paper is properly
installed.
- Close the printer door.
1144 Check that
printer door is
closed and that
paper is present
Printer hardware error.
- Ensure that the printer
paper is properly
installed.
- Close the printer door.
1145 A printer error
has occurred.
Call service
technician
Printer hardware error.
- Ensure that the printer
paper is properly
installed.
- Close the printer door.
1146 Printer paper
replaced
This message is shown in the
Activity Log after
replacement of printer paper.
- No action required. For
information only.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...