ABL90 FLEX reference manual
1. Setup
1-27
Step Action
1.
Press Next slot to display the desired slot and then press Edit.
2.
Select the parameter to be edited, using Next param. or Prev
param.
3.
Press Update to change the range to the one shown in the "Lot to
Date range (2 SD)" box (if available).
4.
Press the check button to activate or deactivate the Fixed SD (i.e. a
minimum allowed control range is defined by setting a Fixed SD).
To change the SD value, touch the "SD" field to highlight it and
enter the value, using the keypad. Confirm the entry with Enter.
5.
Highlight the limit by touching it on the screen and enter your own
value(s), using the keypad. Confirm with Enter.
6.
Repeat the procedure for other parameters in the same manner.
NOTICES:
See I Appendix - Quality control in this manual for detailed
information on statistics and its parameters
"Lot-to-date range (2 SD)" is the range calculated over the
course of the lot, represented mathematically by the mean
value
2 SD; this is the range within which 95 % of the
measurements are found.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...