ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-13
No.
Message
Interpretation
Operator action
613
Cal expired (Cl) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
614
Cal expired (Glu) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 1- or 2-
point calibration.
615
Cal expired (Lac) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as “…..”.
- Perform a calibration.
Removal condition:
- Successful 1- or 2-
point calibration.
616
Cal expired (OXI) Too long time elapsed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 1- or 2-
point calibration.
641
ABL/DMS PC
restarted
The analyzer was restarted
from power off.
- No action required.
For information only.
642
ABL/DMS PC
connected to wet
section
Added by DMS PC when
connection to the wet section
is obtained.
- No action required.
643
ABL/DMS PC
disconnected
from wet section
The connection from the DMS
PC to the wet section is lost.
- Shut down and restart
the analyzer.
- Contact Radiometer
service representative.
648
Calibration failed
or not accepted
The last calibration was
aborted or not accepted.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
- Check for and remedy
system messages.
- Repeat the calibration.
Removal condition:
- Successful calibration.
662
Barometer out of
range
Measured barometer value is
outside the measuring range:
60-106.7 kPa.
- Contact Radiometer
service representative.
669
QC value outside
control range
Measured parameter value is
outside control range.
- Verify the procedure
and repeat
measurement.
- Refer to Quality
Control Systems
Reference Manual.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...