1. Setup
ABL90 FLEX reference manual
1-66
Replacement
schedule reminder(s)
Message on next patient result
Lock analyzer when 10 % overdue
GREEN or
YELLOW
GREEN or
YELLOW
System message(s)
present
Message on next patient result
GREEN or
YELLOW
User activity
reminder(s)
GREEN
or
YELLOW
Built-in QC error(s)
present
Rerun same level once (default OFF)
NOTICE: Critical system messages will always result in a RED traffic light
signal.
The specified traffic light signal and the messages will continue to appear until
the condition no longer exists.
Corrective action
Explanation
"?" on specific
parameters
The affected parameter(s) will be marked with "?" in
subsequent patient results.
Message on next
patient result
The subsequent patient results will be marked on the
Message screen.
Lock analyzer
when QC overdue
If a scheduled quality control measurement is more than
0 % overdue compared with its scheduled time, the
analyzer will be locked.
Lock analyzer
when 10 %
overdue
If a scheduled replacement procedure is more than 10 %
overdue compared with its scheduled time, the analyzer
will be locked.
Explanation of
corrective
actions
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...