ABL90 FLEX reference manual
1. Setup
1-31
Activating Westgard Rules for a specific parameter:
Step Action
1.
Display the desired slot, using Next slot and press Edit.
2.
Display the desired parameter using Next param or Prev param.
3.
Activate the desired Westgard Rule(s) by pressing the
corresponding check button.
(All future quality control data for the given slot/parameter will be
evaluated according to the selected Westgard Rule(s).)
4.
Select Westgard Rules for other parameters or levels in the same
manner.
5.
Press Back to return to the Westgard Rules setup screen.
Selecting/deselecting all Westgard Rules:
Step Action
1.
Display the desired slot, using Next slot.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...