CHAPTER 10:
As a manual conditioning of the sensor cassette pack is not
possible for the time being, references to this option are no longer
valid and, therefore, the following analyzer messages no longer
exist in software version 2.4:
972
Sensor conditioning due
1080
Installed sensor cassette was not registered as conditioned
1333
Sensor conditioning started, no BC
Conditioning in
sensor cassette pack
not possible
1334
Sensor conditioning started, BC
Warm start
With the new warm start feature in software version 2.4 the
following new analyzer message has been included:
No.
Message
Interpretation
Operator action
1340
Sensor cassette
maintenance by
Analyzer has
been interrupted
This message is used in the Activity
log to indicate startup using a sensor
cassette which has been left without
an FTC activity for more than 1 hour.
No action required
TABLE OF CONTENTS/INDEX:
Conditioning in
sensor cassette pack
not possible
As the pack with the sensor cassette cannot be used for
conditioning, the name of the pack has been changed from
Conditioning unit to Sensor cassette pack.
The manual will be updated with the above information as part of the next manual
update.
Please place this note to the users in the binder of your manual.
Technical
documentation
Instructions to
user
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...