6. User-defined corrections
ABL90 FLEX reference manual
6-8
Correction factors for electrolyte and metabolite
parameters
The following corrections to the slope and the offset are possible within the
stated limits for:
Blood samples:
Parameter Slope
Offset
cK
+
0.80-1.20
1.0 mmol/L
cNa
+
0.80-1.20
10 mmol/L
cCa
2+
0.80-1.20
1.00 mmol/L
cCl
–
0.80-1.20
10 mmol/L
cGlu 0.750-1.250
5.0 mmol/L
cLac 0.750-1.250
5.0 mmol/L
The Radiometer default values for the electrolyte and metabolite parameters
must be reset manually by the user on the Parameter setup screen to 1.000
for each parameter.
Correcting
slope and
offset
Resetting
corrections to
default values
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...