5. Sensors and optical system
ABL90 FLEX reference manual
5-18
Measuring principle of the pH and electrolyte sensors
The pH and electrolyte sensors are measured according to the potentiometric
measuring principle, where the potential of an electrode chain recorded at a
voltmeter is related to the concentration of a substance via the Nernst equation.
The electrode chain (or electrical circuit) set up to measure pH/electrolytes is
illustrated in the following diagram:
1
2
3
4
5
6
The electrode chain describes an electrical circuit consisting of the following:
Item Part
Function
1
Voltmeter
Measures the voltage potential in the circuit.
2
Reference electrode
Provides electrical connection to the voltmeter.
3
Liquid junction
Point of contact between the reference sensor
and the sample.
4
Sample
The unknown liquid being measured.
5
Membrane
An ion-sensitive membrane, which is sensitive
to H
+
/electrolyte ions.
6
Solid-state contact
Provides electrical connection to the voltmeter.
Every element in the electrode chain contributes a voltage to the total potential
drop through the chain. Thus:
When immersed in the appropriate electrolyte solution, both electrodes exhibit
separate potentials
The membrane junctions between the sample and electrolyte solutions also
exhibit separate potentials
The total potential across the electrode chain, therefore, is the sum of these
separate potentials, all but one of which are known and constant, as outlined in
the table on next page.
Potentiometric
measuring
principle
Electrode chain
Parts and
description
Electrode chain
potential
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...