10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-30
No.
Message
Interpretation
Operator action
1112 LS inlet not
empty
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1113 LS sensors not
empty
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1114 LS outlet not
empty
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1115 Ws
communication
error: wrong
message format
Internal communication
error.
- Restart the analyzer.
- Contact Radiometer
service representative.
1116 Ws
communication
error: keep alive
timeout
Internal communication
error.
- Restart the analyzer.
- Contact Radiometer
service representative.
1117 Oxi spectrometer
temperature drift
A large deviation in
temperature has been
observed. This is probably
due to a change in the
ambient environment.
- Perform a calibration
1120 Sensor
replacement
successful
This message is shown in the
Activity Log following a
successful replacement of
the sensor cassette.
- No action required. For
information only.
1121 The port did not
open during
sensor
replacement
This message is shown in the
Activity Log after a failed
sensor cassette replacement.
- Reinstall the sensor
cassette.
- Restart the analyzer.
- Contact Radiometer
service representative.
1123 The sensor chip
data could not be
read or written
during
replacement
This message is shown in the
Activity Log after a failed
sensor cassette replacement.
- Reinstall the sensor
cassette.
- Restart the analyzer.
- Contact Radiometer
service representative.
1124 An unregistered
sensor was
installed during
replacement
This message is shown in the
Activity Log after a sensor
cassette replacement that
did not identify a previously
conditioned cassette.
- No action required. For
information only.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...