ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-9
No.
Message
Interpretation
Operator action
378
Calibration
sensitivity out of
range
The sensitivity value is out of
range for the given
parameter.
- Check for and remedy
any system messages.
- Repeat the calibration.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
Removal condition:
- Successful calibration.
379
Calibration
unstable
(response fault)
An electrode response fault
occurred during calibration.
- Check for and remedy
any system messages.
- Repeat the calibration.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
Removal condition:
- Successful calibration.
443
Ca(7.4) not
usable
cCa
2+
at a pH of 7.4 is not
usable as the actual pH is
outside the 7.2-7.6 range.
- No action required.
452
Interference
during
measurement
Interference was detected
during measurement.
- Check the patient
record for medication
containing possible
interfering substances.
484
Today is last day
in stat. month -
remember to
print QC
statistics
After the current day, quality
control statistics obtained
over the month will be
deleted and new statistics
started.
- Print the QC statistics
if a copy is required.
487
A new statistical
month has begun
- remember to
export WDC data
A new statistical month has
begun.
- Make a WDC report
disk.
Removal condition:
- A WDC report disk has
been made.
494
Bilirubin too high Detected bilirubin
concentration, ctBil(blood), is
greater than 2000 µmol/L.
The corresponding plasma
bilirubin concentration can be
calculated as follows:
ctBil(blood) = (1-Hct) ×
ctBil(plasma).
- No action required.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...