ABL90 FLEX reference manual
7. Performance characteristics
7-19
For adult use:
Adult samples within reference range:
The uncertainty in the bilirubin measurement on whole
blood can, in some cases, exceed the level required to
measure normal bilirubin levels for children older than 3
months and adults (bilirubin reference range
4-22 µmol/L).
Adult samples with an increased bilirubin level:
External tests using adult samples were performed on
samples with typically 80 % of the total bilirubin in the
conjugated form. For these highly conjugated samples
the external tests showed a negative bias of 18% on
whole-blood samples.
The patient samples represented typical variations in ctBil, ctHb, sO
2
, pH and MCHC
(Mean Corpuscular Hemoglobin Concentration) values.
Three external tests were carried out at two different sites. Each test had its own ABL90
analyzer – a total of three.
Wet Chemistry analyzer Roche Modular with Roche Calibrator was used as a reference
[Ref. 21]. Each external test site had two Modulars – a total of four. ctBil was measured
in
mol/L.
The field test results are given below:
S
yx
is the spreading around the linear line.
Actual external test from neonatal critical care hospitals using whole blood. Data from
three field tests are merged. Values are in µmol/L.
y = 1.014x - 0.828
R
2
= 0.985
0
50
100
150
200
250
300
350
400
0
100
200
300
400
Modular
AB
L
9
0
Syx=11.6
N=175
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...