1. Setup
ABL90 FLEX reference manual
1-18
Step Action
1.
Highlight the desired item in the "Selected items" box with the
arrow buttons.
2.
Set the default:
If the item has a value, press Keyboard, enter the value and
press Enter on the keyboard to confirm
If the item has a list of options, press List, highlight the option
using the arrow buttons and press Enter
to confirm
3.
Set or change other default values in a similar manner.
NOTICE:
It is not possible to set default values for all items
The values can be changed on a result-by-result basis on the
Patient identification screen
An item placed in the "Selected items" list does not appear in
the "Available items" list
Default values
Changes to
List
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...