ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-15
No.
Message
Interpretation
Operator action
699
Built-in QC
measurement
started due to
calibration error
The analyzer was set up to
perform built-in QC
measurements in case of
calibration errors.
- Check Calibration
Status and remedy any
reported calibration
errors.
700
Scheduled built-
in QC not run
due to errors in
last calibration
Last calibration contained an
error, and the analyzer was
set up to suspend built-in QC
measurements in case of
calibration errors.
- Check Calibration
Status and remedy
calibration errors.
703
QC expired
QC measurement is 25 %
overdue (corrective action
"Lock analyzer" has been
selected in the Setup
program: Corrective
Actions).
- Perform a quality
control measurement.
Removal condition:
- No QC measurements
are pending.
704
Built-in QC
measurement is
repeated
The scheduled QC
measurement was not
accepted; the measurement
was repeated as requested in
the Setup program:
Corrective Actions.
- No action required.
705
Built-in QC
measurement is
repeated twice
The scheduled QC
measurement was not
accepted; the measurement
was repeated twice as
requested in the Setup
program: Corrective Actions.
- No action required.
707
Replacement(s)
overdue by 10
%. Analyzer
locked.
Replacement is overdue by
10 % (corrective action "Lock
analyzer" was selected in the
Setup program: Corrective
Actions). When the analyzer
is locked, scheduled
calibrations are performed,
but no patient samples or QC
measurements are allowed.
- Check Replacement
Status and replace as
required.
- Unlock analyzer in the
Miscellaneous Setup
program.
Removal condition:
- No replacement
pending.
708
Corrective action
not possible due
to empty solution
pack
Scheduled built-in QC
measurement was
requested, but the solution
pack was empty.
- Insert a new solution
pack.
712
FHbF
measurement
not possible
Composition of the blood
sample makes FHbF
measurement too inaccurate,
but OXI parameters are
compensated for HbF.
See explanation in the ABL90
FLEX reference manual.
- If FHbF is wanted
change sample
composition. For
example, elevate sO
2
and tHb.
713
ctBil
measurement
not possible
Blood sample ctHb is so high
that hardly any plasma is left
to measure plasma bilirubin
on.
ctHb > 15.5 mmol/L.
- If ctBil is wanted,
lower the ctHb value.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...