ABL90 FLEX reference manual
1. Setup
1-33
To add a new RiLiBÄK rule, do the following:
Step Action
1.
Press the Add button to display the screen below:
2.
Select the desired parameter from the parameter list shown in the
right side of the screen.
3.
Press
until the first "Lower Limit" box is highlighted and enter
the desired lower limit.
4.
Highlight the next box and select "<" or "<=".
5.
Highlight the first "Upper Limit" box and select "<" or "<=".
6.
Highlight the next "Upper Limit" box and enter the desired lower
limit.
7.
To select the d/- range, press the desired radio button.
8.
Enter the d/- range in the "Ranges" box.
9.
Press Back to return to the RiLiBÄK ranges screen. The added
RiLiBÄK rule is now shown in the screen.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...