10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-20
No.
Message
Interpretation
Operator action
826
HbF locked
The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
827
tBil locked
The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
831
pH unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
832
pCO
2
unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
833
pO
2
unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
834
K unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
835
Na unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
836
Cl unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
837
Ca unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
839
Glu unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
840
Lac unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...