5. Sensors and optical system
ABL90 FLEX reference manual
5-30
Calibration of the pO
2
sensor
The pO
2
sensor is calibrated to determine its sensitivity by measuring one
calibration point during a sensitivity calibration process. Performance of the
sensor from calibration to calibration is checked and during sample or quality
control analysis any drift on sensitivity is checked. The pO
2
sensor is calibrated
on ambient air.
The sensitivity is defined as the percentage of the measured pO
2
on ambient air
compared to the reference value:
2
2
O (meas)
S
O (ref)
p
p
where pO
2
(ref) is the pO
2
tension in ambient air saturated with water vapor:
2
2
2
O (ref)
O (p(amb) pH O)
p
F
where FO
2
is the pO
2
fraction in ambient air, and pH
2
O is the partial water vapor
pressure of saturated air at 37
o
C, and p(amb) is the barometric pressure.
In connection with the sensitivity calibration performed on ambient air, also the
CAL1 (rinse) solution is measured to obtain a status. This status aims to check
the performed calibration. This is done by comparing the measured value of the
CAL1 (rinse) solution to the reference value of CAL1, given by the smart chip):
2
2
2
O (status,cal)
O (CAL1,cal)
O (CAL1,ref)
p
p
p
For every measurement, the pO
2
calibration is checked by comparing the
measured value of CAL1 (rinse) solution to the value obtained on the CAL1
solution of the last calibration (CAL1
CAL
):
2
2
2
O (status,meas)
O (CAL1,meas)
O (CAL1,cal)
p
p
p
Introduction
Sensitivity
Status
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...