7. Performance characteristics
ABL90 FLEX reference manual
7-20
The same data as above but depicted in a Bland-Altman plot below.
Lines indicate “Mean” and +/- “15 umol or 10%”. Values are in µmol/L. Difference =
ABL90 – Modular.
Interference tests
The results from the interference tests are given as the deviation from the correct result
[Ref. 14].
The following interference results are found for the pH and blood gas electrodes:
Interference on ...
Substance
Test
Concentration
pH
(level 7.4)
p
CO
2
p
O
2
mmHg
Test matrix
2% <
0.010
N/A <
1
Blood/Aqueous
Intralipid
5% <
0.010
N/A <
1
Blood/Aqueous
Fluorescein 400
mg/L N/A N/A
<
1
Blood
K
+
17 mM
<
0.010
N/A N/A
Blood
Na
+
190 mM
<
0.010
N/A N/A
Blood
Ca
2+
5.5 mM
<
0.010
N/A N/A
Blood
pH/blood gas
-50
-40
-30
-20
-10
0
10
20
30
40
50
0
100
200
300
400
ctBil
D
iffe
re
n
c
e
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...