ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-5
No.
Message
Interpretation
Operator action
165
LIS/HIS: High-
level protocol
could not
generate high-
level packet
An error occurred while
formatting a message.
- Check protocol
configurations.
Contact Radiometer
service representative.
166
LIS/HIS: General
communication
error
An internal error occurred in
the LIS/HIS communication
module.
- Contact Radiometer
service representative if
the problem persists.
167
LIS/HIS: High-
level protocol
received packet
in wrong format
An error occurred while
parsing (interpreting) a
message.
- Check protocol
configurations.
Contact Radiometer
service representative.
200
User msg:
This is only a message. An
operator has entered a note
in the log.
- No action required.
201
Westgard Rule
(1.2s) violation
Measured parameter value is
outside the mean +/- 2 SD
range.
- Verify procedure and
repeat measurement.
- Check Replacement
Status for pending
replacements.
- Refer to the ABL90
FLEX reference manual
for detailed evaluation
procedure.
202
Westgard Rule
(1.3s) violation
Measured parameter value is
outside the mean +/- 3 SD
range.
- Verify procedure and
repeat measurement.
- Check Replacement
Status for pending
replacements including
electrodes.
- Refer to the ABL90
FLEX reference manual
for detailed evaluation
procedure.
203
Westgard Rule
(2.2s) violation
Two consecutive
measurements are outside
the mean +/- 2 SD range on
the same side of the mean.
This may indicate a shift.
- Verify procedure and
repeat measurement.
- Check Replacement
Status for pending
replacements including
electrodes.
- Refer to the ABL90
FLEX reference manual
for detailed evaluation
procedure.
204
Westgard Rule
(R.4s) violation
The difference between two
consecutive measurements
exceeds 4 SD. This may
indicate an inconsistency in
your procedure or an
unstable analyzer.
- Verify procedure and
repeat measurement.
- Check Replacement
Status for pending
electrode replacements.
- Refer to the ABL90
FLEX reference manual
for detailed evaluation
procedure.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...