ABL90 FLEX reference manual
5. Sensors and optical system
5-33
Construction of the metabolite sensors
The
cGlu and cLac sensors are three-electrode sensors consisting of an internal
silver/silver chloride reference electrode, a platinum auxiliary electrode, and a
platinum anode. The sensors are covered by a multi-layer membrane bound to
the sensor board.
The membrane consists of four layers:
The biocompatible layer
The outer membrane – permeable to
cGlu/cLac
The enzyme layer
The inner membrane – permeable to H
2
O
2
Item Part
Description
1 Biocompatible
layer
Biocompatible layer
2
Outer membrane
Outer membrane permeable to glucose –
diffusion control
3
Enzyme layer
Contains glucose/lactate oxidase.
4
Inner membrane
Cellulose acetate.
5 Reference
Ag/AgCl
electrode.
6 Anode
Platinum
electrode.
7 Cathode
Platinum
electrode.
8
Electrode base
The structural platform on which the sensor is
formed.
The zero current is a small background current measured at the electrode when
no
cGlu/cLac is present in a solution. As CAL 1 contains no cGlu/cLac, a baseline
representing the zero current, I
0
as a function of time (I
0
= f(t)), is obtained
from continuous measurements on CAL 1.
This I
0
baseline is obtained as follows:
At the end of a rinse, with CAL 1 in the measuring chamber, the zero
current of the metabolite electrodes is measured periodically
The previous N (N = 8) measurements on the CAL 1 – before a calibration
or a sample measurement starts – are used to obtain a baseline
representing the time function of I
0
Basic
description
Diagram
Parts and
description
Zero current
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...