ABL90 FLEX reference manual
1. Setup
1-15
Reportable ranges
Step Action
1.
Scroll to the desired parameter, using the up/down arrows or the
scroll bar.
2.
Key in the desired lower limit and confirm with Enter on the
keypad.
3.
Key in the upper limit and confirm with Enter on the keypad.
4.
To change the reportable range to the default (primary) setting,
highlight the desired parameter and press Set default.
5.
To change all parameters to the default values, press Set all
default.
Press Continue to change the reportable ranges of all parameters
to the default ones.
Press Cancel to keep the user-defined reportable ranges and return
to the previous screen.
6.
Press Close to exit and confirm the selected settings.
NOTICES:
A reportable range must be smaller than or equal to the range
of indication
Measured parameters show the reportable ranges. Derived
parameters show "....."
See also Calibration verification, chapter 6 in the ABL90 FLEX
operator's manual.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...