1. Setup
ABL90 FLEX reference manual
1-42
Locking/unlocking a parameter:
Step Action
1.
Ensure that the analyzer is not connected to the RADIANCE system,
as parameters can be locked/unlocked from the RADIANCE system.
2.
Highlight a parameter on the screen, using the scroll facilities.
3.
Press Lock/Unlock. (This button is grayed-out if the analyzer is
connected to the RADIANCE system.)
4.
To unlock a parameter, highlight it and press Lock/Unlock. The
traffic light on Analyzer status will change from YELLOW to a color
corresponding to the analyzer's overall status.
NOTICE: A locked parameter will show YELLOW on the parameter bar and will
change the overall analyzer status traffic light on the Analyzer status screen to
YELLOW. The parameter value will be absent from the printout; however, the
locked parameter will be calibrated.
Editing the parameter setup:
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...