5. Sensors and optical system
ABL90 FLEX reference manual
5-46
The measured spectrum is compared to a model spectrum calculated from the
determined concentrations. The difference between these two spectra is called
the residual spectrum. If this residual spectrum is too high, the oximetry module
parameters
ctHb, sO
2
,
FO
2
Hb,
FCOHb, FMetHb, FHHb, FHbF and ctBil will be
flagged with a warning.
In addition, a warning will accompany the results if any of the following
conditions exist:
ctHb <
0.1 mmol/L or
ctHb > 25 mmol/L
FHb(deriv) < –2 % or FHb(deriv) > 102 %
where
FHb(deriv) is defined as sO
2
,
FO
2
Hb,
FCOHb, FMetHb, FHHb
SHb <
2 % or SHb > 10%
Value of Turbidity <
0.5 % or > 5%
Residual
spectrum
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...