10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-14
No.
Message
Interpretation
Operator action
679
Barometer error The measured parameter
may be unreliable due to
barometer error.
- Contact Radiometer
service representative.
682
OXI module not
active
The OXI module is not
responding due to an internal
communication problem, or
the software configuration
does not match the analyzer
type.
- Shut down the
analyzer, using the
Temporary Shutdown
function; then restart it.
- Contact Radiometer
service representative.
Removal condition:
- OXI module ready, or
software configured
without OXI module
support.
688
ctHb/ceHb too
low for OXI
calculation
ctHb < 1 mmol/L, or ceHb <
0.75 mmol/L. If ctHb is too
low, FHHb, FO
2
Hb, FCOHb
and FMetHb are not
calculated. If ceHb = cHHb +
cO
2
Hb is too low, sO
2
is not
calculated.
- If Oxi derivates are
wanted, elevate tHb
and/or sO
2
.
692
ABL not
connected to
RADIANCE
The analyzer is not
connected to RADIANCE.
- Contact your
RADIANCE/IT engineer.
- Check RADIANCE
Communication Setup
including TCP/IP
address, port no. and
password.
- Check that RADIANCE
is responding.
- Check network
connections.
Removal condition:
- RADIANCE connection
established or disabled.
693
ABL not
connected to
RADIANCE -
incorrect
password
The analyzer was refused
connection to RADIANCE due
to incorrect password.
- Enter the correct
password in the
analyzer's RADIANCE
Communication Setup.
Removal condition:
- RADIANCE connection
established or disabled.
694
ABL connected to
RADIANCE
The analyzer is connected to
RADIANCE.
- No action required.
695
ABL disconnected
from RADIANCE
The analyzer was
disconnected from
RADIANCE.
- No action required.
696
ABL<>RADIANCE
communication
error
Communication error
between the analyzer and
RADIANCE.
- Contact Radiometer
service representative.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...