ABL90 FLEX reference manual
5. Sensors and optical system
5-17
Construction of the pH and electrolyte sensors
The pH and electrolyte sensors are of solid-state design with a H
+
, K
+,
Na
+
and
Ca
2+
sensitive PVC membrane. The Cl
–
sensor is of solid-state design with a Cl
–
sensitive epoxy membrane.
The pH sensor is used as an example:
3
1
4
1
2
3
4
Item Part
Description
1
Membrane
Ion-selective membrane that is in direct
contact with the sample or calibration
solution and that is sensitive to a specific ion,
e.g. the H
+
ions.
2 Solid-state
contact
The point of electrical and ionic contact with
the membrane.
3
Electrical contact The point of electrical contact between the
sensor and the analyzer.
4
Electrode base
The structural platform on which the
electrode is formed.
Diagram
Parts and
description
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...