Appendix – Quality control
ABL90 FLEX reference manual
I-14
This rule is a rejection rule.
The difference between
two consecutive
measurements exceeds 4
SD.
This indicates inconsistency in your procedures or
an unstable analyzer.
Corrective action:
Perform steps described in section Quality control
evaluation
further in this appendix.
This rule is applicable to both manual and built-in QC.
This rule is a rejection rule.
Four consecutive
measurements outside
the mean
1 SD on the
same side of the mean.
This indicates a trend or shift.
Corrective action:
Perform steps described in section Quality control
evaluation
further in this appendix.
This rule is only applicable to manual QC and is only recommended when
user-defined control ranges are established.
This rule is a rejection rule.
10 consecutive
measurements on the
same side of the mean.
This indicates a trend or shift.
Corrective action:
Perform steps described in section Quality control
evaluation
further in this appendix.
This rule is only applicable to manual QC and is only recommended when
user-defined control ranges are established.
Rule R
4s
4
1s
Rule 10
x
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...