8. Parameters
ABL90 FLEX reference manual
8-6
Measured parameters
The following is the used:
m =
male
f =
female
Reference
range
For adult arterial blood
Reference
[10]
(unless otherwise specified)
Definition
Indicates the acidity or alkalinity of the sample.
Unit -
Range of indication
6.300-8.000
Reference range
7.35-7.45 (m, f)
Definition
Concentration of hydrogen ions in blood.
Unit nmol/L
Range of indication
10.0-501
Reference range
35.5-44.7 (m, f)
Is used both for blood and expired-air samples.
Definition
Partial pressure (or tension) of carbon dioxide in
blood.
High and low pCO
2
values of arterial blood
indicate blood hypercapnia and hypocapnia,
respectively.
Unit mmHg;
kPa;
Torr
Range of indication
mmHg; Torr: 5.0-250
kPa: 0.67-33.3
Reference range
mmHg: 35-48 (m); 32-45 (f)
kPa: 4.67-6.40 (m); 4.27-6.00 (f)
Conversion of units
p(kPa) = 0.133322 × p(mmHg) = 0.133322 ×
p(Torr)
p(mmHg) = p(Torr) = 7.500638 × p(kPa)
General
information
pH
cH
+
pCO
2
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...