ABL90 FLEX reference manual
1. Setup
1-69
A message, sent from the RADIANCE system to the connected analyzer and
displayed on the main screen, can be changed or deleted in this program.
Step Action
1.
Press the Keyboard button, type the message (up to 40 characters
long) and confirm with Enter.
To delete the current message, press Delete on the keyboard, or
delete a message and type a new one, if desired.
2.
Confirm the change with Enter to return to the Miscellaneous
setup screen.
Step Action
1.
Check that the “Enable screen saver” check button is activated in
the “Setup” box.
2.
In the “Screen saver” box, select the time with the arrow buttons.
Analyzer
messages
Setting the
time for the
screen saver to
appear
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...