ABL90 FLEX reference manual
5. Sensors and optical system
5-5
General information
Calibration is the process that relates the electrode signals during the calibration
sequence to the values of the calibrating solutions and air. Calibration enables
the electrode signals to be converted to the accurate values for an unknown
sample.
Calibration must be performed at regular intervals so that normal variations in
sensor output can be compensated for after inevitable minor changes in the
sensor's behavior.
Calibration of all sensors is performed using air, CAL 1 (also used for rinse), CAL
2 and CAL 3 (see chapter 9, Solutions for more information on the solutions).
The calibration solutions contain known concentrations of the substrates to be
measured. These concentrations are vital in determining the measurement
accuracy of the analyzer.
The concentration of each substance in the calibration solutions is programmed
into the integrated smart chip of the solution pack. The information is
automatically read by the analyzer when a solution pack is installed in the
analyzer.
The traceability certificate for the solution pack is found in chapter 9 of this
manual.
Definition
Frequency
Calibration
solutions
Traceability of
calibration
solutions
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...