ABL90 FLEX reference manual
5. Sensors and optical system
5-23
Construction of the pCO
2
sensor
Item Part
Description
1
Silicone membrane A membrane separating the sample and the
electrolyte solution. Is only permeable to CO
2
.
2
Electrolyte solution A solution separating the silicone membrane and
the pH membrane, Ag/AgCl sensors. The
electrolyte solution is vented by CO
2
whereby pH
is changed.
3 pH
membrane H
+
sensitive membrane.
4 Reference
Ag/AgCl
electrode
5 Solid-state
contact
for the pH system
The point of electrical contact between the pH
membrane and the analyzer.
6 Solid-state
contact
for the Ag/AgCl
system
The point of electrical contact between the
reference electrode and the analyzer.
7
Electrode base
The structural platform on which the electrode is
formed.
Diagram
Parts and
description
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...