1. Setup
ABL90 FLEX reference manual
1-30
Westgard Rules setup
In this program you can select Westgard Rules for all slots or for specific
parameters.
Step Action
1.
Select the desired slot, using Next slot.
2.
Press On/Off to activate the assigned Westgard Rules for the slot
or press this button again to deactivate them.
NOTICE: The Westgard Rules are a set of statistical rules. When applied to the
QC results, they can increase the probability of detecting an error in the
sampling procedure or in the analyzer itself, or they can help detect a shift or
trend in your QC results by comparing current measurement values of a QC
solution with previous values.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...