1. Setup
ABL90 FLEX reference manual
1-14
Step Action
1.
Highlight a parameter on the Reference ranges and critical
limits screen, using the up/down arrows.
2.
Enter sample type, sex and age group, if required.
3.
Press Edit to edit entries for a highlighted parameter.
4.
If any entries are present and you cannot use any of them, press
Clear limits.
Then enter new critical and reference limits, using the keypad and
confirming each entry with Enter.
5.
To change a value, touch and highlight it. Then enter the limit and
confirm with Enter.
6.
Press Back to return to the Reference ranges and critical limits
screen.
7.
Highlight a parameter in the "Parameter list" box to view the limits
on the Reference ranges and critical limits screen.
Setting
reference and
critical limits
for each
parameter
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...