ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-21
No.
Message
Interpretation
Operator action
841
tHb unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
842
MetHb unlocked The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
843
COHb unlocked The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
844
HHb unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
845
O
2
Hb unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
846
sO
2
unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
847
HbF unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
848
tBil unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
852
RADIANCE:
Message from RADIANCE.
- No action required. For
information only.
855
Base Excess out
of range
Base Excess exceeds the +/-
30 mmol/L range.
- For information only.
No analyzer error was
detected.
875
Sample aged
The specified limit for sample
age has been exceeded.
- Draw and analyze new
sample.
885
Cyclic QC
schedule reset
from RADIANCE
The cyclic QC schedule has
been reset and all related
reminders have been
removed as a result of a
RADIANCE command.
- No action required. For
information only.
886
LIS/HIS: No valid
POCT1A DML
Device ID file
A file with a valid Device ID
does not exist. A valid Device
ID is needed in order to use
the POCT1A DML protocol.
- Contact Radiometer
service representative
to obtain a Device ID
file.
Removal condition:
- Valid Device ID found.
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...