ABL90 FLEX reference manual
6. User-defined corrections
6-5
Correction factors for oximetry parameters
The following corrections can be user-defined for the oximetry parameters:
Parameter Allowed
user-defined
corrections
Slope
Offset
ctHb Yes No
sO
2
Yes Yes
FCOHb No
Yes
FMetHb No
Yes
FO
2
Hb No
No
FHHb No No
FHbF
Yes Yes
ctBil
Yes Yes
NOTICE: In order to define the corrections accurately the measurements of the
oximetry parameters on the ABL90 FLEX analyzers should be made without any
entered corrections. To avoid truncation errors from an enabled "Out-of-range-
suppression" function it is important to disable the function.
The following recommendations apply to the oximetry parameters:
Item Description
Units
g/dL; g/L; mmol/L
Sample
Set ctHb of a SAT100 sample to
15 g/dL (9.3 mmol/L) and
pH
7.4
ctHb, maximum
point
Uncorrected or corrected:
15 g/dL or 9.3 mmol/L
Slope
0.950-1.050
Item Description
Units Fraction
Sample
Set ctHb of gas equilibrated SAT0 and SAT100 samples to
15 g/dL (9.3 mmol/L) and pH
7.4
Slope
0.900-1.100
Offset
0.050
Allowed
corrections
ctHb
sO
2
Summary of Contents for FLEX ABL90
Page 1: ...ABL90 FLEX reference manual ...
Page 2: ......
Page 3: ...ABL90 FLEX reference manual ...
Page 4: ......
Page 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Page 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Page 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Page 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Page 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Page 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Page 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Page 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Page 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Page 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Page 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...