Appendices . 149
Appendix 4: Fluid Path Materials
Ensure that any sample or cleaning materials loaded into the flow cell are compatible with
the ZnSe flow cell windows and the materials in the fluid path (Table 6).
CAUTION
Never use strong acids or bases, or oxygen-containing solvents such as
aldehydes, ketones or ethers, with the system without first checking
compatibility, as these liquids may cause immediate and severe damage.
Ethyl and isopropyl alcohol may be used safely.
Table 6 Materials in fluid path in contact with samples
System Component
Material
Injection Station Fill Cups
Polyether ether ketone
Injection Station Male to Female Luer
Connector
Polypropylene
Flow Cell / Injection Station Tubing
Material
Fluorinated ethylene propylene
Red Luer-lok Connector
Ethylene tetrafluoroethylene
Blue Nut Connector
Polypropylene
Ferrule
Ethylene tetrafluoroethylene
Fluid Control Module Internal Tubing
Viton®
Waste Liquid Container (L9004517)
Bottle
Cap
Polypropylene
Polypropylene
4-oz Bottles (N9308900)
Acrylonitrile butadiene styrene
17-mm Test Tubes (N0777167)
Polypropylene
Flow Cell
ZnSe windows