Maintenance and Troubleshooting . 119
Checking the Waste Liquid Container
The waste liquid container should be emptied regularly. Dispose of all liquid waste according
to accepted laboratory practices.
We recommend that you use a waste container that is larger than the system liquid
container. If you routinely empty the waste container whenever you refill the system liquid
container, the waste container will not overflow.
NOTE: The system liquid container available with the OilExpress 4 system is 2 L
(part number L9004516) and the waste container is 5 L (part number L9004517).
Cleaning the VersaTips
Check the VersaTips daily for dirt or damage. The VersaTips are cleaned when the
instrument is primed. Refer to
on page 74.
NOTE: If a tip is damaged, contact your PerkinElmer Service Representative.
Weekly Preventative Maintenance
Priming the System
If the system has been unused for any extended period, you should Prime the system using
the control on the Autosampler Control dialog. This is done to remove any residue, deposits
on page 74.
At the end of the prime, check the fluid lines from the system liquid container to the
VersaTips for any air bubbles. You may need to prime the system two or three times before
you run an analysis.
Summary of Contents for OilExpress 4
Page 1: ...OilExpress 4 Oil Condition Monitoring System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 17: ...Warnings and Safety Information ...
Page 27: ...Overview of the OilExpress 4 System ...
Page 67: ...Getting Started with OilExpress 4 ...
Page 95: ...Maintenance and Troubleshooting ...
Page 144: ...144 OilExpress 4 System User s Guide ...
Page 145: ...Appendices ...