Maintenance and Troubleshooting . 131
System liquid tubing contains
air after priming (continued)
System liquid tubing has a leak
or a blockage
Inspect lines for blockages. Tighten
tubing and connections to the
syringes
on page 120
If you are unable to resolve the
issue, contact your PerkinElmer
Service Representative
Syringe to VersaTip fluid
lines contain air after priming
System not primed after
maintenance
Use the Prime button on the
Autosampler Control dialog to
Prime system 2–3 times more (until
there is no air in the tubing) before
on page 74
Syringe or tubing has a leak or a
blockage
Inspect lines for blockages. Tighten
tubing and connections to the
syringes
on page 120
If you are unable to resolve the
issue, contact your PerkinElmer
Service Representative
System liquid tubing is not
immersed in system liquid
Check the system liquid tubing is in
the liquid. Refill system liquid, if
required
Use the Prime button on the
Autosampler Control dialog to
Prime system 2–3 times more (until
there is no air in the tubing) before
on page 74
Shuttle error
Sample shuttle is not connected
to mains power
Connect sample shuttle to splitter
cable from Spectrum Two. Check
that the spectrometer is connected
to the mains power and that the
green LED is lit
on page 46
Sample shuttle obstructed
Remove any obstruction
Check the hinged arm is in the
correct position
on page 97
Sample shuttle hardware failure
Contact your PerkinElmer Service
Representative
Summary of Contents for OilExpress 4
Page 1: ...OilExpress 4 Oil Condition Monitoring System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 17: ...Warnings and Safety Information ...
Page 27: ...Overview of the OilExpress 4 System ...
Page 67: ...Getting Started with OilExpress 4 ...
Page 95: ...Maintenance and Troubleshooting ...
Page 144: ...144 OilExpress 4 System User s Guide ...
Page 145: ...Appendices ...