136 . OilExpress 4 System User's Guide
Liquid is not dispensed
during a flush (continued)
Syringe or tubing has a leak or a
blockage
Click
Stop Analysis
Inspect lines for blockages. Tighten
tubing and connections to the
syringes
on page 120
If you are unable to resolve the
issue, contact your PerkinElmer
Service Representative
System is not primed
Click
Stop Analysis
Use the Prime button on the
Autosampler Control dialog to
Prime system 2–3 times more (until
there is no air in the tubing) before
on page 74
Liquid welling out of fill cup
Waste container has become
pressurized
Waste tubing is in the waste liquid
Release pressure
Empty the waste container, if
required
Fill cup blocked
Clean fill cup
FCM is not connected to the
correct fill cup
Check that the tubing from the fill
cup on the injection station is
connected to the flow cell, and that
the tubing from the flow cell is
connected to the FCM
End of Analysis
Sample not detected by
liquid presence sensor
Sample was not aspirated
(Sample was not in correct
position on autosampler deck)
Check that all sample racks are
located correctly on the deck for
the current system configuration
Sample was not aspirated
(Actual height of sample in
container does not correspond to
sampling height in the software)
Check the sampling heights on the
Autosampler Setup dialog, and in
the Samples screen (if specified)
Problems with sample
preparation:
Sample is outside viscosity
range supported by the system
Sample contains large particles
Check the viscosity of the sample.
Ensure that the sample was filtered
to remove any large particles
Summary of Contents for OilExpress 4
Page 1: ...OilExpress 4 Oil Condition Monitoring System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 17: ...Warnings and Safety Information ...
Page 27: ...Overview of the OilExpress 4 System ...
Page 67: ...Getting Started with OilExpress 4 ...
Page 95: ...Maintenance and Troubleshooting ...
Page 144: ...144 OilExpress 4 System User s Guide ...
Page 145: ...Appendices ...